国产精品青草久久福利不卡,国产精品毛片无遮挡高清,久久香蕉国产线看观看6,久久久久久老熟妇熟女

本站熱搜:氧化鋯氧量分析儀,盤裝式氧量分析儀,電磁流量計,WSS雙金屬溫度計
  • 技術文章ARTICLE

    您當前的位置:首頁 > 技術文章 > Common faults and causes of zirconia oxygen analyzers

    Common faults and causes of zirconia oxygen analyzers

    發(fā)布時間: 2024-12-13  點擊次數(shù): 315次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


產(chǎn)品中心 Products
在線客服 聯(lián)系方式 二維碼

服務熱線

0550-2409223

掃一掃,關注我們

皖公網(wǎng)安備 34118102000850號

精品国产污免费网站入口| 丁香六月婷婷综合缴情欧美| 国产性夜夜春夜夜爽夜夜| 日本不卡久久伊人麻豆传媒| 国产三级电影精品麻豆| 亚洲国产精品久久久久网站| 国产一区二区综合资源| 911国产精品视频| 97久久人人超碰国产精品| 国产综合精品久久久久成人蜜臀| 日韩精品高清在线观看| 国产一二三区亚洲欧美| 亚洲国产精品成人综合色在线| 精品人妻一区二区三区20| 精品国产片亚洲一区| 亚洲人妻御姐中文字幕| 国产精品内射久久久久| 久久久区一区二区三| 盗盗摄婷婷精品一区二区| 久久久久蜜桃成人精品一区| 蜜臀91精品国产高清在线| 亚洲精品欧美日韩专区| 欧美亚洲一区二区成人| 欧美日韩精品一区二区三区激情在线| 欧美激情a成人综合亚洲综合| 久久久区一区二区三| 国产精品亚洲欧美日韩一区在线| 欧美日本大陆一区二区| 久久久久久99国产精品免费| 花花草草寻亲记哪里看全集| 亚成区一区二区人妻熟女| 日本欧美一区二区免费不卡| 欧美日韩一区二区中文字幕| 日韩一区二区三区久久久| 在线电影日韩一区二区三区| 欧美成人伦理在线播放| 久久久久久久久久福利高潮| 韩国三级电影视频网站| 欧美做爰猛烈大尺度| 国产一区二区三区欧美日| 欧美日韩国产综合在线视频|